What does 3M recommend regarding load and pack monitoring with biological indicators, Class 5 integrating indicators, and Class 6 emulating indicators?
3M recommendations for load monitoring:
- To monitor to the highest level of assurance, 3M considers Every Load Monitoring with a BI to be a best practice.
- Daily and implant loads, preferably every load monitoring, using a Process Challenge Device (PCD) containing a biological indicator.
- A PCD containing a BI and Class 5 integrating indicator in every load containing implants.
- A class 5 integrating indicator may be used within a PCD to monitor non-implant sterilizer loads.
3M recommendations for pack monitoring:
- To monitor to the highest level of assurance use one or more Class 5 Integrating Indicators inside each package or container.
- Following AAMI ST79 recommended practice by using a Class 4 or Class 5 CI inside each package. (Section 10.5.2.2.2 Internal Chemical Indicators)
- A Class 6 Emulating Indicator could be used as an internal indicator at the pack/tray level in cycles for which it is labeled.