3M™ Comply™ (SteriGage™) Steam and EO Chemical Integrator Test Packs


3M™ Comply™ (SteriGage™) Steam and EO Chemical Integrator Test Packs
Click to Enlarge

These test packs present a challenging situation within the sterilizer to monitor its performance. A sophisticated Class 5 integrating chemical indicator inside a porous PCD pack shows that effective sterilant penetration has occurred.

The chemical integrators inside these test packs are Class 5 integrating indicators. 3M™ Comply™ Steam and EO Chemical Integrators can also be used as the process challenge device (PCD) to release processed items, excluding implants (load control), however, their use does not replace the use of biological indicators in PCDs.

The 3M™ Comply™ Steam Chemical Integrator Test Pack consists of a 3M™ Comply™ Steam Chemical Integrator inside a porous substrate contained in a moisture impervious disposable wrap. The 3M™ Comply™ EO Chemical Integrator Test Pack consists of a Comply EO Chemical Integrator inside a syringe barrel along with moisture absorbent material. The barrel is enclosed with thermal insulating material and sealed inside a peel pouch.


Image Catalog # Product Description MSDS
Comply Steri-Gage Chemical Integrators 41360 and 1243   41360   3M™ Comply™ (SteriGage™) Steam Chemical Integrator Test Pack 41360  Integrator Test Pack with Comply™ (SteriGage™) steam chemical integrator  PDF  
Benefits

  • Pack design provides a consistent, resistant challenge to the sterilization cycle for reliable results.
  • Integrators respond to all steam (i.e., time, temperature, and saturated steam) or EO sterilization critical parameters (i.e., EO concentration, time, temperature and relative humidity) to assure accuracy.
  • Test packs provide immediate Accept/Reject results to allow you to respond more quickly to sterilizer problems.
  • Results are revealed at a glance for quick and easy reading.
  • Can provide early readout of equipment performance prior to biological indicator results and/or in between biological indicator-monitored loads for faster detection of sterilization process failures.

    Suggested Applications

  • To monitor sterilizer performance.
  • Use Comply™ steam chemical integrator test pack for 250°F/121°C gravity and 270°F/132°C vacuum-assisted steam sterilization processes.
  • For steam sterilization a dark color should enter the accept window.
  • Use Comply™ EO chemical integrator test pack for EO/HCFC gas mixture sterilization process.
  • For EO sterilization a blue color should enter the accept window.